Assessment of fracture behaviour in heat treated NiTi rotary endodontic files following multiple uses: An in-vitro study
Aim and Objectives: Evaluate the fracture behaviour and warning signs (e.g., deformation) of two heat?treated nickel–titanium rotary files—EdgeEndo and TorqueEndo—during repeated simulated clinical use, and determine whether fracture risk differs between them.
Introduction: Nickel-titanium (NiTi) rotary endodontic files have revolutionized root canal therapy due to their superior flexibility and shape memory. However, despite advancements in heat treatment technologies that enhance their mechanical properties, these instruments remain susceptible to fracture, especially after repeated clinical use. Understanding the fracture behaviour of heat-treated NiTi files is essential for improving their safety and durability. This in-vitro study aims to assess how multiple uses influence the fracture resistance and failure modes of these instruments under simulated clinical conditions.
Materials and Methods: This in?vitro pilot study used 16 extracted maxillary first premolars with ~5° curvature (n?=?8 per file type). Each file was used up to 16 times or until fracture, following crown?down preparation, standard irrigation, and sterilization protocol. Files were inspected at 10× magnification after each use for permanent deformation or micro cracks. Fisher’s Exact Test was applied for statistical analysis due to the small sample size (??=?0.05).
Results: Edge Endo: 5 fractures; 4 preceded by bending, Torque Endo: 6 fractures; all preceded by deformation. No significant difference in fracture frequency was found (p?=?0.64). Warning signs typically appeared after approximately six uses.
Conclusion: Both file systems displayed visible warning signs—primarily bending—prior to fracture. No conclusive difference in fracture risk was observed, although this pilot study is limited by size, canal allocation bias, and its in?vitro setup. These findings support further large-scale, randomized studies with microstructural documentation.
Keywords: Cyclic fatigue resistance, File separation, Heat treatment, Microstructural analysis
References
- Chakka NVMK, Ratnakar P, Das S, Bagchi A, Sudhir S, Anumula L. Do Niti instruments show defects before separation? Defects caused by torsional fatigue in hand and rotary nickel-titanium (niti) instruments which lead to failure during clinical use. J Contemp Dent Pract. 2012;13(6):867–72.
- Inan U, Gonulol N. Deformation and fracture of Mtwo rotary nickel- titanium instruments after clinical use. J Endod. 2009;35(10):1396-
- Sotokawa T. An analysis of clinical breakage of root canal instruments. J Endod. 1988;14(2):75–82.
- Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26(3):161-5.
- Kwak SW, Shen Y, Liu H, Wang Z, Kim HC, Haapasalo M. Heat treatment and surface treatment of nickel–titanium endodontic instruments. Front Dent Med. 2021; 2:769977.
- Shen Y, Zhou HM, Zheng YF, Campbell L, Peng B, Haapasalo M. Metallurgical characterization of controlled memory wire nickel- titanium rotary instruments. J Endod. 2011;37(11):1566–71.
- Zhou HM, Shen Y, Zheng W, Li L, Zheng YF, Haapasalo M. Mechanical properties of controlled memory and superelastic NiTi wires used in the manufacture of rotary endodontic instruments. J Endod. 2012;38(11):1535–40.
- Hou X, Yahata Y, Hayashi Y, Ebihara A, Hanawa T, Suda H. Phase transformation behaviour and bending property of twisted nickel- titanium endodontic instruments. Int Endod J. 2011;44(3):253–8.
- Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41(7):1134–8.
- Al-Amidi AH, Al-Gharrawi HAR. Effect of autoclave sterilization on the cyclic fatigue resistance of EdgeFile X7, 2Shape, and F-one nickel–titanium endodontic instruments. J Conserv Dent. 2023;26(1):26–30.
How to Cite This Article
Vancouver
Narayanan P, SV R, Simon EP, Raveendran C, Koroth S. Assessment of fracture behaviour in heat treated NiTi rotary endodontic files following multiple uses: An in-vitro study [Internet]. IP Indian J Conserv Endod. 2025 [cited 2025 Oct 02];10(2):109-112. Available from: https://doi.org/10.18231/j.ijce.2025.019
APA
Narayanan, P., SV, R., Simon, E. P., Raveendran, C., Koroth, S. (2025). Assessment of fracture behaviour in heat treated NiTi rotary endodontic files following multiple uses: An in-vitro study. IP Indian J Conserv Endod, 10(2), 109-112. https://doi.org/10.18231/j.ijce.2025.019
MLA
Narayanan, Parvathy, SV, Ravi, Simon, Elsy P, Raveendran, Chandini, Koroth, Sarin. "Assessment of fracture behaviour in heat treated NiTi rotary endodontic files following multiple uses: An in-vitro study." IP Indian J Conserv Endod, vol. 10, no. 2, 2025, pp. 109-112. https://doi.org/10.18231/j.ijce.2025.019
Chicago
Narayanan, P., SV, R., Simon, E. P., Raveendran, C., Koroth, S.. "Assessment of fracture behaviour in heat treated NiTi rotary endodontic files following multiple uses: An in-vitro study." IP Indian J Conserv Endod 10, no. 2 (2025): 109-112. https://doi.org/10.18231/j.ijce.2025.019