Get Permission Bukhari, Bhat, and Rashid: Impact of nickel-titanium instrument heat treatment on the precision of an inbuilt electronic apex locator and endodontic motor


Introduction

Determining the working length (WL) is crucial for the outcome of the endodontic procedure since the greatest results are obtained when the apical limit is precisely determined and set up near where there is apical constriction. The process of root canal preparation and obturation necessitates strict confinement within the root canal system.1 The cement dentinal junction is widely regarded as the optimal end point.2 In practical clinical settings, however, pinpointing this histological feature is unfeasible. As a result, the apical constriction is deemed an acceptable reference for canal preparation. Determining the appropriate working length (WL) involves employing both radiographic techniques and electronic apex locator in daily practice.

Despite its common use, the radiographic approach's accuracy has been challenged by several studies.3, 4, 5 This method relies on two-dimensional imagery and the subjective interpretation of the clinician, often leading to potential over- or under instrumentation. 5, 6, 7, 8, 9 Additionally, many root canals exhibit foramens located on the lateral root surfaces, limiting accurate radiographic WL determination. 3

Suzuki originally briefed about EALs for measuring root canal lengths during 1942, 4 and Sunada reportedly employed them in clinical settings in 1965.5 Although some authors claim that the anatomical diameter of the apical foramen,10 instrument's tip diameter, and type of alloy might result in interference in the precision of these instruments in estimating root canal length. 11, 12, 13

Manufacturers have developed various nickel-titanium (NiTi) alloys with heat and surface treatments to enhance the mechanical properties of NiTi instruments, resulting in increased flexibility and resistance to torsional and cyclic fatigue. 13, 14, 15 However, the potential effects of these treatments on the electrical circuit impedance remain uncertain.

The WL and BMP can now be done simultaneously thanks to various endodontic motors that feature an integrated EAL. The concurrent use of EAL is preferred because root canals with significant curvatures may experience alterations in work length during the shaping phase. 16

Since an integrated EAL and motor unit is employed during BMP, the current research sought to assess ex vivo the effects of various thermal processes on NiTi equipment as well as the diameter of the apical preparation on an EAL's accuracy.

Materials and Methods

Preparing the teeth

In the current research, 20 extracted human Maxillary incisor with completely formed apex in each group a single canal verified by periapical radiograph, and the curvature ranging from 10° to 30° were obtained using schneider’s method. These teeth were formerly standardized at 17 mm root length by removing dental crowns with a diamond disc.

Establishment of the control working length

WL was assessed via a manual stainless steel K-file #15 (Dentsply Maillefer, Ballaigues, Switzerland) and a clinical microscope (Zeiss, Oberkochen, Germany). To identify the WL, the file was pushed into the root canal up until it could be seen flush with the main foramen.

Measurement of working length electronically

The roots were then secured within an acrylic container and covered with a conductive gel that contained 0.5 percent KCl, and a saltwater solution and 2.5 percent hydroxyethyl cellulose. 6 

WL was electronically measured using a J. Morita Endomotor TRI AUTO ZX2 (Frankfurt, Germany) that has an inbuilt EAL.

The study utilized NiTi systems of tip diameter of 0.25 and various heat treatments: HyFlex CM size 25.04 (Coltene, Altstätten, Switzerland), Twisted file Adaptive size 25.06 (Axis/SybronEndo, Orange, CA, USA), WaveOne Gold Primary size 25.,07 (DentsplyMaillefer, Ballaigues, Switzerland), Reciproc R25 size 25.08 (VDW, Munich, Germany), and Reciproc Blue R25 size 25.08 (VDW, Munich, Germany).

Root canals were flushed with 15 mL of 2.5% sodium hypochlorite, and electronic working length (WL) was ascertained when the file reached the point where the orange LED light became visible on the endodontic motor screen. Subsequently, the silicone stopper was adjusted at the occlusal reference, and the length was measured using an endodontic gauge (Dentsply Maillefer, Ballaigues, Switzerland). Three measurements were taken for each instrument.

If the readings were within 0.5 millimeters of each other, and the measuring was deemed to be accurate (scoring 0), although discrepancies more than 0.5 mm were deemed incorrect (score 1).

Statistical analysis

The significance level was set at (P<0.05) and IBM SPSS statistics software was used for the data analysis. To compare the results collected through comparing the various heat treatments or tip diameters, Fisher's exact test was employed to assess the variances between and between groups.

Results

There wasn't any significant variation in EAL accuracy in measuring working length amongst different groups (P > 0.05). Instruments with various thermal treatments and an identical 0.25 mm tip diameter did not result in any statistically significant differences in EAL accuracy (Table 1).

Table 1

Accurate and inaccurate results obtained using 0.25 mm diameter instruments with different heat treatments.

Difference ≦ 0.5 mm

Difference > 0.5 mm

Mtow 25

20

0

TF 25

20

0

wave one gold primary

19

1

R25

19

1

R25 blue

19

1

[i] *indicate statistical difference between group (P<0.05)

Discussion

The purpose of the research was to assess the impact of heat treatment and endodontic instrument tip width on the precision of an EAL integrated into an endodontic motor. According to the findings of this investigation, the thermal evaluation of treatments (M-wire, R-phase, Gold Wire, Blue Analysis of wire, CM wire, and tip diameters (0.25 mm,0.35", 0.45", 0.45" and 0.50") revealed no influences on the EAL's accuracy during BMP. The null hypothesis was accepted.

The accurate calculation of the WL is one of the parameters that determines whether endodontic therapy will be successful.3 The WL, particularly in curved canal, may vary throughout the BMP.16, 17, 18 It is advantageous to follow the WL during root canal preparation when endodontic motors and EAL are used paired together.19, 20, 21 The precision of EALs is frequently evaluated in the scientific literature using ex vivo investigations.1, 10, 11, 21 To assess the correctness of EALs, Duran-Sindreu et al.22 compared in vivo and ex vivo models and discovered no differences. Ex vivo studies can also more precisely control sample variability and conditions for experimentation because they are straightforward, reproducible, and standardized.11, 21 We used a conventional, easily produced 2.5% hydroxyethyl cellulose gel that conducts electricity.11 Sodium hypochlorite, as in earlier studies,1, 11, 21, 23 was utilized as an irrigation solution to mimic clinical use.

In order to calculate WL utilizing the differential between resistance and capacitance (impedance) inside the root canal, EALs function as electricity conductors in the electrical circuit.23 Whenever an endodontic equipment approaches the apical constriction, impedance values change and are converted into WL measurements in millimeters from the apical foramen.12, 23, 24 The composition of NiTi alloys changes with different thermal treatments, affecting the percentage of different crystalline phases (martensite and austenite), which affects the mechanical characteristics of the equipment.15

The study's findings indicate that the precision of the EAL device was unaffected by the thermal processes of the NiTi alloy (P = 0.17). The crystalline structure of NiTi can also affect electrical resistivity values given that the martensite phase's resistivity is 14.7% more than the austenite phase's. Compared to other alloys, CM wire contains a higher proportion of martensite-phase crystals in its crystalline structure.13 Although CM alloys had higher resistivity, there were no appreciable variations in the WL readings.

Conclusion

It was discerned that the variations induced by the diverse heat treatment approaches did not result in any statistically significant differences. It is important to acknowledge the inherent limitations of the study, which encompass factors such as sample size, experimental conditions, and the specific heat treatment protocols applied. However, within the confines of these limitations, the results consistently pointed towards a lack of substantial divergence in the material's properties and precision of apex locator.

Source of Funding

None.

Conflicts of Interest

There are no conflicts of interest.

References

1 

D Ricucci K Langeland Apical limit of root canal instrumentation and obturation, part 2. A histological studyInt Endod J199831639440910.1046/j.1365-2591.1998.00183.x.

2 

K Suzuki Experimental study on iontophoresisJapan J Stomatol19421641129

3 

MJ Altenburger Y Cenik JF Schirrmeister KT Wrbas E Hellwig Combination of apex locator and endodontic motor for continuous length control during root canal treatmentInt Endod J20094243687410.1111/j.1365-2591.2008.01535.x

4 

MP Gordon NP Chandler Electronic apex locatorsInt Endod J20043774253710.1111/j.1365-2591.2004.00835.x

5 

I Sunada New method for measuring the length of the root canalJ Dent Res19624135787

6 

E Akisue G Gavini JA De Figueiredo Influence of pulp vitality on length determination by using the Elements Diagnostic Unit and Apex LocatorOral Surg Oral Med Oral Pathol Oral Radiol Endod200710441293210.1016/j.tripleo.2007.04.018

7 

E Akisue SD Gratieri FB Barletta CL Caldeira R Grazziotin-Soares G Gavini Not all electronic foramen locators are accurate in teeth with enlarged apical foramina: an in vitro comparison of 5 brandsJ Endod20144011091210.1016/j.joen.2013.09.032

8 

V Chopra S Grover S D Prasad In vitro evaluation of the accuracy of two electronic apex locatorsJ Conserv Dent200811282510.4103/0972-0707.44056

9 

S Y Abidi M Azfar T Nayab A Shaukat M Hasan NN Baig Accuracy of working length measurement with endo motor having built-in apex locator and comparison with periapical radiographsJ Pak Med Assoc202070343744110.5455/JPMA.302464

10 

P M Gehlot V Manjunath M K Manjunath An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand filesRestor Dent Endod201641161110.5395/rde.2016.41.1.6

11 

L Cianconi V Angotti R Felici G Conte M Mancini Accuracy of three electronic apex locators compared with digital radiography: An ex vivo studyJ Endod201036122003710.1016/j.joen.2010.08.036

12 

U Aydin E Karataslioglu F Aksoy C Yildirim In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diametersJ Conserv Dent201518166910.4103/0972-0707.148899

13 

G Gavini MD Santos CL Caldeira ME Machado LG Freire EF Iglecias Nickel-titanium instruments in endodontics: A concise review of the state of the artBraz Oral Res201832suppl 16710.1590/1807-3107bor-2018.vol32.0067

14 

E Pedullà GRML Rosa C Virgillito E Rapisarda HC Kim L Generali Cyclic Fatigue Resistance of Nickel-titanium Rotary Instruments according to the Angle of File Access and Radius of Root CanalJ Endod2020463431610.1016/j.joen.2019.11.015

15 

S Staffoli N M Grande G Plotino T Özyürek M Gündoğar L Fortunato Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel-titanium rotary instrumentOdontology20191073301710.1007/s10266-018-0399-5

16 

V Aggarwal M Singla S S Bhasin Influence of instrument size and varying electrical resistance of root canal instruments on accuracy of three electronic root canal length measurement devicesInt Endod J20175055061110.1111/iej.12649

17 

SW Schneider A comparison of canal preparations in straight and curved root canalsOral Surg Oral Med Oral Pathol1971322271510.1016/0030-4220(71)90230-1

18 

J L Caldwell Change in working length following instrumentation of molar canalsOral Surg Oral Med Oral Pathol1976411114810.1016/0030-4220(76)90260-7

19 

M Tien H Tjoa M Zhou PV Abbott Comparative study of four endodontic file systems to assess changes in working length during root canal instrumentation and the effect of canal curvature on working length changeJ Endod20204611101510.1016/j.joen.2019.10.004

20 

DW Christofzik A Bartols M Khaled B Grosner-Schreiber CE Dörfer The accuracy of the auto-stop function of different endodontic devices in detecting the apical constrictionBMC Oral Health20171714110.1186/s12903-017-0425-y

21 

MM Ali R Wigler S Lin AY Kaufman An ex vivo comparison of working length determination by three electronic root canal length measurement devices integrated into endodontic rotary motorsClin Oral Investig20162082303810.1007/s00784-016-1903-3

22 

F Duran-Sindreu S Gomes E Stöber M Mercadé L Jané M Roig In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigantsInt Endod J20134687697410.1111/iej.12057

23 

MH Nekoofar MM Ghandi SJ Hayes PM Dummer The fundamental operating principles of electronic root canal length measurement devicesInt Endod J200639859560910.1111/j.1365-2591.2006.01131.x

24 

BC Vasconcelos LM Bastos AS Oliveira RA Bernardes MA Duarte N Vivacqua-Gomes Changes in root canal length determined during mechanical preparation stages and their relationship with the accuracy of root ZX IIJ Endod201642111683610.1016/j.joen.2016.07.022



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.

  • Article highlights
  • Article tables
  • Article images

Article History

Received : 24-06-2023

Accepted : 26-08-2023


View Article

PDF File   Full Text Article


Copyright permission

Get article permission for commercial use

Downlaod

PDF File   XML File   ePub File


Digital Object Identifier (DOI)

Article DOI

https://doi.org/10.18231/j.ijce.2023.029


Article Metrics






Article Access statistics

Viewed: 781

PDF Downloaded: 232